• A dialog comes up in the shell
4. To check and eventually download a new firmware (from T
OPSPIN)
•type 2→ ENTER
The check starts and ends for example as shown in the following picture
if there is a problem with a component you get an information to contact your
service office
if you need a firmware update you will get a message (A) with the length of time
for the download (B). It is recommended to do the download as soon as possible.
5. Press y → ENTER to start the download
During the download it is absolutely necessary that spectrometer and
workstation are not disturbed! Make sure that no one pulls a cable out of
the spectrometer or will crash the Linux system while using it in parallel!
(Yes we know this seems very improbable, but if the download would
crash, your board might be unusable afterwards)
• A ’normal’ error during downloading the new firmware (you will see a com-
ment in the shell) requires to do the download once again!