62 Configuration of TOPSPIN
DONE
INDEX
INDEX
software. If you have to click on Cf or Expinstall depends on what you want to do
with T
OPSPIN. You are mainly in one of the following situations:
1. You want to configure a processing-only-PC with a default configuration on a
datastation → go to chapter 2.3.1 (for details see chapter 6.2.3)
2. You have installed the new software on the same PC as a previous X
WIN-
NMR/TOPSPIN installation, so the old spectrometer configuration is available on
the same PC → go to chapter 2.3.2 (for details see chapter 6.2.1)
3. You want to configure a processing-only-PC like your spectrometer-PC and
you have the spectrometer configuration on another computer → go to chapter
2.3.3 (for details see chapter 6.2.2)
4. You have to configure the software for controlling a spectrometer on a new
hard disk without a backup of the spectrometer configuration, so you have no
configuration files → go to chapter 6.2.4
6.2.1 If your PC controls a spectrometer
1. Be sure that the spectrometer is booted (see chapter 9.3.2 and 12.3.1.7)
2. Open a dataset
•Type new and create a new dataset.
• You can also select a dataset (e.g. exam1d_1H’) acquired with another
T
OPSPIN or XWIN-NMR version.
3. Enter cf (The command cf now configures the software according to the
hardware of your spectrometer):
a) A window will appear:
Password request. Please enter the NMR ADMINISTRATION
PASSWORD
Type in the NMR ADMINISTRATION PASSWORD. Click OK.