Configuration of TOPSPIN 65
DONE
INDEX
INDEX
nected. The RS232/485 channel for external devices-
Table appears.
Check if the tty-numbers of the hardware components are set correctly.
f) A window will appear:
Additional configuration. Security check. Enable peak
power check (POWCHK). Enable puls power check (PUL-
CHK).
Choose the options if you need this security options.
g) Only if you have a sample changer! A window will appear:
Should the Sample Changer control the Lift? (yes/no)
The answer depends on how the air flow is connected, normally you will
answer: yes ENTER
In this cases the question appear:
Delay between SX and next command [sec]?
Type in the length of the delay e.g. 10 ENTER
h) The
nuclei-table appear.
Check if the frequency of the nuclei are set correctly. If in doubt you can
use default values by clicking on Restore then click Save
i) The results of cf appears in a text window. Check this list carefully and
compare the result with an list from an earlier configuration (if available)
click Print and store the paper output with your other spectrometer
documentation, then click OK. These data are saved in
<TS_HOME>/conf/instr/<spect>/uxnmr.info.
If during ’cf’ an error message occurs that invites you to do a firmware
update see chapter 6.3
After finishing cf the window configure will appear showing a list of configu-
ration steps. Accept the default selection and click Start (You will be asked for
!